Once-A-Day® Ecos™ Bowie-Dick Test is a compact eco-friendly, lead free method of daily sterilizer testing.
- Compact disposable Bowie-Dick test pack
- Lead and heavy metal free ink
- Distinct blue to pink color change
Description
Once-A-Day® Ecos™ Bowie-Dick Test Pack is a compact, fully disposable, and environmentally friendly Bowie-Dick test pack. It is lead, heavy-metal, and latex-free and is completely biodegradable. The test sheets and inks contain no lead or heavy metals, no toxic concerns, pollutants, latex, dry natural rubber, or health hazards.
Benefits
- Eco-friendly: Once-A-Day® Ecos™ test pack is a more sustainable option than traditional Bowie-Dick test packs, as it is completely biodegradable and does not contain any harmful chemicals.
- Compact size:Once-A-Day® Ecos™ test pack is compact, saving space in your inventory.
- Easy to use: Once-A-Day® Ecos™ test pack is easy to use and interpret, making it ideal for busy healthcare settings.
- Reliable: Once-A-Day® Ecos™ test pack is manufactured under strict quality control standards and is compliant with AAMI / ISO 11140-1:2010, Class 2 and AAMI / ISO 11140-5:2007 standards.
Applications
The Once-A-Day® Ecos™ test pack is designed for pre-vacuum sterilizer mechanical air removal testing at the start of each day, or after routine or other maintenance. It can detect low temperature/pressure conditions, wet, and superheated steam.
How to Use
For complete guide to use follow the IFU.
- Place the Once-A-Day® Ecos™ test pack inside the sterilizer chamber on the bottom rack.
- Run a standard pre-vacuum sterilization cycle at 273.2°F/134°C using a 3.5-minute exposure time.
- After the sterilization cycle is complete, remove the Once-A-Day® Ecos™ test pack from the sterilizer chamber and examine the test sheet.
Interpreting the Results
A properly processed Once-A-Day® Ecos™ test pack will exhibit a uniform black color change across the entire surface of the sheet. Any deviation from a full color change, such as white areas or uneven coloring, indicates a sterilizer malfunction.
The environmentally friendly benefits include test sheets and inks that contain no lead or heavy metals, no toxic concerns, pollutants, latex, dry natural rubber, or health hazards.
AAMI / ISO 11140-1:2010, Class 2
AAMI / ISO 11140-5:2007
For use at 273.2°F/134°C using a 3.5 min exposure time
Storage Conditions: 15°-32°C (59°-90°F) 30-70% Relative Humidity