ONIX™ Helix Chemical Indicators

ONIX™ Helix Chemical Indicators is dedicated for hollow loads in a bench top sterilizer.

  • Helix type Bowie & Dick Test for steam sterilizers at 134°C for 3.5 minutes
  • Stable white to black color change that can be maintained for record keeping
  • Multi-use Device lasts 250 cycles

Not for sale in USA.

SKU: 70000100 Categories: ,

Description

Propper ONIX™ Helix is dedicated for hollow loads in a bench top sterilizer. This Bowie & Dick test stimulates a hollow instrument and generates a bigger challenge for the steam sterilization process.
The reusable nature of Propper ONIX™ eliminates excess waste that is commonly found in Bowie & Dick test packs.

Propper ONIX™ Helix test device is dedicated for hollow loads in a bench top sterilizer. This test stimulates challenge of hollow instruments and generates a bigger challenge for the steam sterilization process.
The reusable nature of Propper ONIX™ eliminates excess waste that is commonly found in paper process challenge packs and reduce cost of testing.

ONIX™ chemical indicator strips are designed for use with HT™ Helix PCD Test devices which are parts of 26942100 and 26942200 HT™ Helix process challenge test kits. The purpose of the test is load control. The test is equally suitable for monitoring hollow and porous loads. ONIX™ chemical indicator strips can be universally used in challenged devices for small and large size sterilizer chambers. Distinct color change from off-white to black allows for easy result interpretation. After processing, stable non-reverting ink guarantees reliable record keeping for 25 years. ISO 11140-1:2014, Type 4. Strip works in the temperature range of 121°C-135°C.

ONIX™ Bowie & Dick chemical indicator strips are designed for use with HT™ Helix Bowie & Dick Test devices. ONIX™ Bowie & Dick chemical indicator strips have distinct color change from off-white to black which allows for easy result interpretation. Indicator strips show a fail result in presence of air and low temperature. After processing, stable non-reverting ink guarantees reliable record-keeping for 25 years. ISO 11140-1:2014, Type 4. ONIX™ chemical indicator strips are designed for use with HT Helix & Bowie Dick type Test in steam sterilizers operating at 134°C for 3.5 minutes.

Propper ONIX Helix PCD Device Small Chamber
SKU: 26910900
Each

Propper ONIX Helix Hollow Load PCD Strip
SKU: 26911300
1 Pack of 250

Propper ONIX Helix Bowie Dick Indicator Strip
SKU: 26911400
1 Pack of 250

Ordering

Literature

Below you will find associated information about this product.