26923100 - BI-OK Plasma SCBI Vials

26923100 - BI-OK Plasma SCBI Vials
Category: Sterilization

Quick Overview

BI-O.K. Plasma Self-Contained Biological Indicators are intended for use in routine monitoring of VH2O2 sterilizer cycles with or without a plasma phase.


BI-O.K. Plasma Indicators combine a vaporized hydrogen peroxide compatible carrier with Geobacillus stearothermophilus spores (1.0-5.0 x 106) and culture media for ease-of-use and the high level of assurance of a biological indicator result. The label contains a process indicator and can be written on for record keeping purposes and vial identification.


Process the vial in a vaporized hydrogen peroxide sterilization cycle, crush glass ampule and incubate in the Propper BI-O.K. incubator at 55-60oC for 24 hours before observing the result. Vials are compatible with standard size incubators. 


Sterilization failure is indicated by observing a yellow color. A red/orange color signifies no spore growth and is your indicator that sterilization conditions have been met. 


ANSI/AAMI/ISO 11138-1 compliant

Availability: In stock

26923100 - BI-OK Plasma SCBI Vials
Category: Sterilization

Quick Overview

BI-O.K. Plasma Self-Contained Biological Indicators are intended for use in routine monitoring of VH2O2 sterilizer cycles with or without a plasma phase.


BI-O.K. Plasma Indicators combine a vaporized hydrogen peroxide compatible carrier with Geobacillus stearothermophilus spores (1.0-5.0 x 106) and culture media for ease-of-use and the high level of assurance of a biological indicator result. The label contains a process indicator and can be written on for record keeping purposes and vial identification.


Process the vial in a vaporized hydrogen peroxide sterilization cycle, crush glass ampule and incubate in the Propper BI-O.K. incubator at 55-60oC for 24 hours before observing the result. Vials are compatible with standard size incubators. 


Sterilization failure is indicated by observing a yellow color. A red/orange color signifies no spore growth and is your indicator that sterilization conditions have been met. 


ANSI/AAMI/ISO 11138-1 compliant

Details

BI-OK Plasma SCBI Vials, 20 per bag